4.9 Operation Status Group
The Operation Status group, consisting of the Operation Condition register, the Operation Positive
Transition register, the Operation Negative Transition register, the Operation Event register, and the
Operation Event Enable register, is used to determine the specific condition that set bit 7 in the
Summary Status Byte. The bits in the Operation Event register provide the following:
4.10 Questionable Status Group
The Questionable Status group, consisting of the Questionable Condition register, the Questionable
Positive Transition register, the Questionable Negative Transition register, the Questionable Event
register, and the Questionable Event Enable register, is used to determine the specific condition that
set bit 3 in the Summary Status Byte.