[SENSe:]FUNCtion?
Query the measurement function and return a quoted string.
[SENSe:]<function>:RANGe {<range>|MINimum|MAXimum}
Select a range for the selected function. For frequency and period
measurements, range applies to the signal’s input voltage, not its
frequency (use FREQuency:VOLTage or PERiod:VOLTage). MIN selects the
lowest range for the selected function. MAX selects the highest range.
[SENSe:]<function>:RANGe? [MINimum|MAXimum]
Query the range for the selected function. For frequency and period, use
FREQuency:VOLTage or PERiod:VOLTage. Autorange thresholds: Down range at
<10% of range; Up range at >120% of range.
[SENSe:]<function>:RANGe:AUTO {OFF|ON}
Disable or enable autorange for the selected function.
[SENSe:]<function>:RANGe:AUTO?
Query the autorange setting. Return “1” (ON) or “0” (OFF).
[SENSe:]<function>:RESolution {<resolution>|MINimum|MAXimum}
Set resolution for the selected function (not valid for frequency, period, or ratio).
Specify the resolution in the same units as the measurement function, not in
number of digits. MIN selects the smallest value accepted, which gives the most
resolution. MAX selects the largest value accepted which gives the least
resolution.
[SENSe:]<function>:RESolution? [MINimum|MAXimum]
Query the resolution for the selected function.
[SENSe:]UNIT {Cel|Far|K}
Select units for temperature measurement. Cel means Celsius, Far means
Fahrenheit,and K means Kelvin.
[SENSe:]UNIT?
Query units for temperature measurement.