A sequence of ASCII-encoded bytes(“SENS”*SCALE+OFFSET) terminated by a new
line and EOI.
Note
It should be the last query in a terminated program message or error-440 will be
generated.
:CALCulate[1]:MATH[:EXPRession]:NAME SCALE_OFFSET
Or
:CALCulate[1]:MATH[:EXPRession]:SELect SCALE_OFFSET
Defines the name of the expression used in mathematic operation.
Return
A sequence of ASCII-encoded bytes:SCALE_OFFSET.
:CALCulate[1]:MATH:STATe <Boolean>
Enables or disables the mathematic operation.
Return
An ASCII-encoded byte. 0 means OFF, and 1 means ON.
Note
This command will clear the limit counts(:CALC:LIM:FCO, :CALC:LIM:PCO).
The command :CALCulate2
This group performs limit testing and data transfer.
:CALCulate2:FEED “[:]CALCulate[1]”
Sets or queries the data flow to be fed into the CALCulate2 block.
Return
The string “CALC” is returned.
:CALCulate2:IMMediate
Instructs the counter to re-compute the current data without re-acquiring data.
This command will not affect: