If the current measurement function is Totalize or Voltage Peaks, a “0” will be
returned and error -221 will be generated.
:CALCulate2:LIMit:FCOunt:LOWer?
Queries the number of the measurement that fails the lower limit test.
Return
Numeric data transferred as ASCII bytes in <NR1> format.
Notes
If the limit test is off, a “0” will be returned and error -221 will be generated.
If there is no valid measurement result, a “0” will be returned and error -230 will be
generated.
If the current measurement function is Totalize or Voltage Peaks, a “0” will be
returned and error -221 will be generated.
:CALCulate2:LIMit:FCOunt[:TOTal]?
Queries the total number of the measurement that fails the limit test.
Return
Numeric data transferred as ASCII bytes in <NR1> format.
Notes
If the limit test is off, a “0” will be returned and error -221 will be generated.
If there is no valid measurement result, a “0” will be returned and error -230 will be
generated.
If the current measurement function is Totalize or Voltage Peaks, a “0” will be
returned and error -221 will be generated.
:CALCulate2:LIMit:FCOunt:UPPer?
Queries the number of the measurement that fails the upper limit test.
Return
Numeric data transferred as ASCII bytes in <NR1> format.
Notes
If the limit test is off, a “0” will be returned and error -221 will be generated.
If there is no valid measurement result, a “0” will be returned and error -230 will be