Voltage range -5.125V to 5.125V for ×1 attenuation
-51.25V to 51,25V for ×10 attenuation
Voltage resolution
0.0025V for ×1 attenuation
0.025V for ×10 attenuation
<source list> default (@1)
range (@1)
:MEASure[:SCALar][:VOLTage]:PERiod? [<expected_value> [,<resolution>]]
[,(@1)|(@2)|(@3)]
Sets the measurement arming to digits and measures the period.
The <expected value> and <resolution> parameters are used to configure the number of digits
of resolution arming setting.
The Channel 1 and Channel 2 share the same trigger setting so that the measurement channel
has auto-trigger enabled at 50% with a positive slope.
<expected value>
Ch1, Ch2
Default 100 ns
Range
2 ns ~ 10 sec
Resolution
<expected value> should be within 10% of the input
period for optimum arming configuration
<expected value>
Ch3
Default 2 ns
Range 0.17 ns ~ 2.67 ns
Resolution
<expected value> should be within 10% of the input
period for optimum arming configuration
<resolution>
Ch1, Ch2
Default Value which indicates 4 digits of resolution for the
specified <expected value>
Range
1E-23 ~ 1E-2 sec (15 to 3 digits of resolution)
Resolution
<resolution> should use a mantissa of 1.0 and be an
even power of 10
<resolution>
Ch3
Default Value which indicates 4 digits of resolution for the
specified <expected value>
Range
1E-24 ~ 1E-11 sec (15 to 3 digits of resolution)
Resolution
<resolution> should use a mantissa of 1.0 and be an
even power of 10
<source list> Default (@1)
Range (@1)(@2)(@3)
:MEASure[:SCALar][:VOLTage]:PHASe? [(@1),(@2)]
Sets the measurement arming to auto” and measures Phase.