Bnc 1105 Universal Counter Instrukcja Użytkownika Strona 147

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 197
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 146
147
147147
147
front panel gating settings and the interface command parameters, please refer
to the TINT table in Chapter 5.4.
[:SENSe]:TINTerval:ARM:ESTART:LAYer2:SLOPe POSitive NEGative
Sets or queries the slope of the external start arm signal used in external arming Time
interval measurements.
Return
A sequence of ASCII-encoded bytes. POS means Positive and NEG means Negative.
Notes
This will be set to Positive after the counter has been turned off or a remote interface
reset.
This command only applies when [:SENSe]:TINT:ARM:ESTART:LAY2:SOUR EXT is
used.
[:SENSe]:TINTerval:ARM:ESTARt:LAYer2:SOURce IMMediate EXTerna
Sets or queries the start arm for Time interval measurements.
Return
A sequence of ASCII-encoded bytes. IMM means Immediate and EXT means
External.
Note
This will be set to Immediate after the counter has been turned off or a remote
interface reset.
[:SENSe]:TINTerval:ARM:ESTARt[:LAYer[1]]:ECOunt <numeric value>
Sets or queries the number of events used to delay the start arm for Time Interval
measurements. The range is from 1 to 99,999,999. The resolution is 1.
Return
Numeric data transferred as ASCII bytes in <NR1> format.
Notes
This will be set to 1 after the counter has been turned off or a remote interface reset.
This command only applies when [:SENSe]:TINT:ARM:ESTART[LAY[1]]:SOUR
Przeglądanie stron 146
1 2 ... 142 143 144 145 146 147 148 149 150 151 152 ... 196 197

Komentarze do niniejszej Instrukcji

Brak uwag